How to solve the problem of refrigeration pump circulator pump failure?
Cooling heating devices are required in many chemical pharmaceutical and laboratory applications. Then how to solve the failure of the refrigeration heating device is also concerned about how to solve it?
1. Circulating pump failure
The circulating pump fault is generally the thermal overload of the circulating pump. There are four main reasons for this fault: The first is that the outer circulating pipeline of the heat conducting oil is too long. At this time, it is necessary to shorten the outer circulation pipeline, reduce the bending of the outer circulation pipeline, and then click Equipment electrical cabinet surface thermal relay Reset button; the second is the circulation pipeline blockage, cooling and heating device which is good to tell everyone, check the circulation pipeline cleaning, install the filter, and then click the equipment electrical cabinet surface thermal relay Reset button; The third is that the circulating pump voltage is too low. It is recommended to measure whether the circulating pump voltage meets the specified range, improve the voltage environment, click the Reset button of the equipment electrical cabinet surface thermal relay; the fourth is that the thermal relay regulates the current too small. At this time, it is recommended to measure the pump current. Compare the range of thermal relays, Phillips screwdriver to properly rotate the thermal relay current, click on the equipment electrical cabinet surface thermal relay Reset button.
2. Sensor failure
A great possibility of sensor failure is wiring error. Once the fault is encountered, check the wiring situation. At this time, you should pay attention to two situations, one is sensor short circuit or temperature <-160 °C, and the other is sensor open circuit or temperature > 349 ° C.
If the sensor is short-circuited or the temperature is <-160 °C, it may be the fault of the sensor itself. At this time, it is recommended to replace the sensor (there is a spare sensor in the equipment slot). If the sensor is wired incorrectly, It is recommended to check the wiring (corrected circuit diagram) correction.

If the sensor is open circuit or the temperature is >349°C, it should be the fault of the sensor itself. It is necessary to replace the sensor in time (there is a spare sensor in the equipment slot). At this time, it is necessary to pay attention to whether the temperature is too high and the temperature is too high. Check whether the temperature is exceeded. Range, if the equipment is not powered, pay attention to check the power supply of the equipment.
These two faults of the cooling heating device are still relatively simple. I believe that everyone who has seen the sharing can solve it very quickly.
Recommandations connexes
-
Avantages du réchauffeur d'huile à transfert de chaleur LNEYA
1028Le réchauffeur électrique à huile caloporteuse est un nouveau type de dispositif de conversion de l'énergie thermique qui utilise l'électricité comme source d'énergie et l'huile caloporteuse comme fluide caloporteur. Il peut fournir de l'huile de transfert de chaleur à la pompe de circulation...
Voir les détails -
Cooling and heating of the reactor and the principle of use
1221Cooling and heating circulator thermostats are commonly used with glass reactors for biochemical instruments. They are widely used in modern fine chemicals, biopharmaceuticals, scientific research and experiment industries. They can be concentrate...
Voir les détails -
13p semi-closed air-cooled compression condensing unit failure warning instructions
1232The 13p semi-closed air-cooled compression condensing unit is used in industrial cold processing. After using it for a long time, some problems will arise, Wuxi Guanya (LNEYA) reminds. There is usually a high-pressure alarm on the 13p semi-closed ...
Voir les détails -
Méthode de refroidissement de la chambre d'essai de choc thermique à trois éléments intégrés
1281The three integrated thermal shock test chambers are designed and manufactured according to user requirements. They are suitable for high and low temperature impact tests of high-temperature or low-temperature environments and for high-temperature...
Voir les détails