How To Choose the Ultra Low Temperature Recirculating Chiller Expansion Valve?
Selection and installation of the expansion valve for refroidisseurs à recirculation à ultra basse température: the selection of expansion valve plays an important role in the performance of the ultra low temperature recirculating chiller refrigeration system. The correct choice of expansion valve will maximize the use of the evaporator and make the evaporator always be matched to the heat load.
First, the selection of thermal expansion valve.
In the ultra low temperature recirculating chiller refrigeration system, the selection and installation of the expansion valve is very important. Only the correct selection and installation can ensure the normal operation of the unit. Therefore, when selecting the expansion valve, we can use the appropriate size of the thermal expansion valve according to the size of the evaporator load and the evaporating temperature. The capacity of the commonly used expansion valve is about 20% larger than the evaporator heat load. For small chiller refrigeration systems, gas-filled thermal expansion valves are generally used, and medium-sized ultra-low temperature circulating chiller refrigeration systems mostly use adsorption-filled or mixed-filled expansion valves.
Second, the expansion valve installation.
1. The chiller expansion valve is installed in front of the evaporator and is usually installed in a vertical position. Considering the ease of installation and maintenance, the cooling loss is small and should be as close as possible to the inlet of the evaporator.
2. The working condition of the expansion valve for the ultra low temperature recirculating chiller has a direct relationship with the installation position of the temperature sensing package. Because the expansion valve completely adjusts the opening size of the expansion valve by the temperature change of the temperature sensing package. The temperature sensor package is installed on the pipe at the outlet of the chiller evaporator. As the evaporator outlet pipe grows, its superheat gradually increases. Therefore, the position of the expansion valve temperature sensing package is directly related to the amount of fluid supplied from the expansion valve.
Recommandations connexes
-
Système de simulation de refroidissement pour l'essai de performance extrême d'un bloc-batterie
1157With the development of technology, pure electric vehicles are becoming more and more common, but pure electric vehicles are difficult to get a better solution in terms of cruising range, charging speed, and pollution after battery waste. This mak...
Voir les détails -
Equipment attention points of laboratory high and low temperature cycle equipment manufacturers
1091In the laboratory, a wide range of experimental instruments are widely used. When the experimental instruments are in operation, they are inseparable from the high-low temperature circulating device refrigeration heating equipment. What should be ...
Voir les détails -
Conditions environnementales pour l'installation de refroidisseurs à air
1105Comme nous le savons tous, l'installation de la plupart des machines doit être adaptée à l'environnement et aux conditions. Bien entendu, nos refroidisseurs industriels à condensation par air ne font pas exception. Si le refroidisseur à condensation par air n'est pas installé pour la première fois, il apportera beaucoup de ...
Voir les détails -
Effect Of Humidity On Laboratory Chillers
1043The high humidity reduces the insulation strength of the electrical equipment of the chiller in the laboratory. The moisture in the air adheres to the surface of the insulating material to form condensation, which reduces the insulation resistance...
Voir les détails