Several Common Troubleshooting Methods for Ultra-low Temperature Freezer in Laboratories
Although the refrigeration system and control circuit of ultra low temperature freezer in laboratory are complex, the common faults are mainly circulating system and circuit part.
Following are several common faults of ultra-low temperature refrigerators and their troubleshooting methods:
1. After power supply is switched on, the air switch of the main power supply is automatically disconnected.
Elimination method: Check whether the compressor of refrigeration system is damaged or not, and replace the parts if damaged. After replacing the fittings, turn on the machine to check whether it is normal. The fault is usually caused by unstable power supply, because the * stage starts first, so the damage is usually in stage. In this case, as long as the delay regulated power supply is installed, the fault will not occur again.
2. Stop working before the temperature in the box reaches the set temperature.
Exclusion: Although the temperature in the box has not reached the set temperature, the panel display temperature has reached the set temperature. This is due to the damage of the thermistor which detects the temperature and the need to replace the parts. If there is no original accessories, a set of temperature control system can be re-installed to eliminate the failure.
3. Poor refrigeration effect may be refrigerant leakage
4. If the ultra-low temperature refrigerator does not start in winter, it may be that the thermostat regulation value is too small.
We are a ultra low temperature freezer manufacturer. Welcome to choose our products.
Recommandations connexes
-
How does the cooling and heating system achieve simultaneous temperature control?
1038Temperature control system The thermostat has heating and cooling functions, and can use the same medium for continuous temperature control from -120℃~320℃. The heating and cooling system of cooling and heating system is mainly composed of four pa...
Voir les détails -
Comment prolonger la durée de vie de l'huile caloporteuse dans un système de contrôle dynamique de la température ?
991Dans l'industrie chimique, le fluide caloporteur de l'huile caloporteuse du système de régulation dynamique de la température du réacteur est principalement déterminé par l'huile caloporteuse elle-même, la température de fonctionnement de l'huile caloporteuse, la dyna...
Voir les détails -
Description du prix de l'équipement Vocs et de l'effet de condensation
1150En tant que fabricant d'équipements vocaux, de nombreux utilisateurs accordent une attention particulière au prix des équipements vocaux lors de l'achat de ces derniers. Que savez-vous donc du prix et de l'effet de condensation des équipements vocaux et de l'importance de la performance ? Du...
Voir les détails -
Integrated control system high and low temperature cycle device ZLF series use items
1067With the development of society, the chemical industry has been further developed, and the temperature system control of the reactor has been strengthened, so as to ensure that the monitoring environment is in a good state, to ensure that the temp...
Voir les détails